is also a mul-drop mode tha
t allows for the connecon of more than one nEXT pump to
a single serial port on the control system.
5.6.1 Connect the serial interface to the customer control equipment
The serial interface is available in RS485 or RS232 opons which can be selected using
the slide switch located adjacent to the main power lead.
To adjust the slide switch, toggle the slide switch using a small tool. The motor controller
default seng is RS232 serial interface. Be sure to replace the round seal to ensure the
IP rang of the pump is sll achieved.
CAUTION:
When connecng the pump t
o a PC, note that the 0 V pin on the RS232 connector
may be connected to earth through the PC. If this is the case, ensure that the 0 V rail
of the 48 V d.c. supply is not also connected to earth at some other point such as at
the power supply. If the 0 V rail of the 48 V d.c. supply will not be connected to earth
at the PC, an opto-isolated interface to the PC should be used.
The nEXT pump can connect directly to the RS485 or RS232 serial input on the control
equipment or a PC as shown in Figure: Logic interface connecons - RS232 serial control
and Figure: Logic interface connecons - RS485 serial control. In this conguraon the PC
is the serial link master and the nEXT pump is the slave. The distance over which the
serial link will work is dependent on any dierence in voltage between the 0 V at the
sending and receiving end. If the 0 V reference at the receiving end is within 0.3 V of the
0 V Control Reference pin on the nEXT pump control connector, then the serial link
should be capable of operang at distances up to 6 m. An interface circuit external to
the nEXT pump may be required for longer distances.
The soware in the nEXT pump is capable of operang with several pumps connected to
a single serial link master. This is referred to as mul-drop mode. The RS485 opon is
recommended for mul-drop mode. With the RS232 opon selected, some addional
hardware will be required to link several nEXT pump units to a single serial link master. A
concept drawing of one possible arrangement is shown in Figure: Conceptual diagram
for mul-drop connecon using RS232 interface. Conversely when the RS485 opon is
selected, connecng several nEXT pumps to a single master becomes a simple wiring
exercise as shown in Figure: RS485 mul-drop connecon.
08/2021 - ©Edwards Limited
Page 60B8J200880_D
300812671_002_C3
B8J200880_D - Inst