5.7 Connecon for mix
ed parallel and serial operaon
The pump can be controlled using parallel interface control inputs, and at the same me
monitor various pump parameters using the serial interface or the USB service port
using the Edwards nST2 PC soware. Alternavely, the pump can be controlled using
commands sent over the serial interface while at the same me monitoring the normal
signal and analogue output over the parallel interface.
Figure: Logic interface connecon - mixed parallel and serial operaon shows a
schemac diagram of a system that demonstrates how to do this. This connecon is a
hybrid of the parallel and serial connecon which are described in detail in Connect the
parallel control and monitoring on page 58 and Connecon for serial control and
monitoring on page 59 respecvely. Many of the opons described in those secons are
available in mixed parallel and serial operaon but note that whilst serial enable is acve
to enable the serial link, the parallel standby and fail signals are not available. The mul-
drop connecon shown in Figure: Logic interface connecon - mixed parallel and serial
operaon can also be used with mixed parallel and serial operaon.
The pump cannot be controlled using both the parallel and serial interfaces
simultaneously. For example, if the pump is started by sending a start command over the
serial interface, the pump cannot then be stopped by using the start /stop switch on the
parallel interface. The pump will ignore the state of the start / stop switch on the parallel
interface. To stop the pump, send a serial stop command. Only when the serial stop
command has been received by the pump can any commands sent via the parallel
interface be acted on.
Similarly, if the pump is started using the start switch on the parallel interface, the pump
cannot then be stopped by sending a stop command over the serial interface. The pump
will ignore any start or stop commands received over the serial interface. To stop the
pump, use the parallel stop switch. Only when the pump has been stopped using the
parallel interface switch will any start or stop commands be accepted via the serial
interface.
08/2021 - ©Edwards Limited
Page 62B8J200880_D
300812671_002_C3
B8J200880_D - Inst
allaon

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards nEXT730D and is the answer not in the manual?

Edwards nEXT730D Specifications

General IconGeneral
BrandEdwards
ModelnEXT730D
CategoryWater Pump
LanguageEnglish

Related product manuals