▪ Send the pump to an Edwards Service Centre for a bearing replacement.
▪ Purchase a bearing service kit and replace the bearing on-site (refer to Bearing on-
site main
tenance on page 106).
8.2 Rotor life
The fague life of nEXT Turbomolecular pump rotors is typically 40,000 to 50,000 cycles.
As a precauonary measure, Edwards recommends that pumps are returned for a major
service (rotor replacement) aer 20,000 cycles of acceleraon to full speed and back to
a stop, or aer ten years of use, whichever occurs rst.
8.3 Cleaning the pump
WARNING:
Clean the ext
ernal surfaces of the pump in a well venlated locaon. When using
cleaning soluons and solvents to clean the pump, observe all precauons specied
by the manufacturer. Avoid inhalaon of any parculates which may be present in
the pump.
NOTICE:
Do not aempt t
o clean any parts of the pump other than external surfaces. Organic
solvents may damage internal pump components. Do not use abrasive materials to
clean any part of the pump.
If the inside of the pump is contaminated, it may not be possible to achieve the specied
ulmat
e vacuum performance, or pump-down mes may increase.
In these circumstances the pump must be returned to an Edwards Service Centre where
the pump will be dismantled and cleaned. Refer to Return the equipment or components
for service on page 105.
Any organic solvent can be used to clean the external surfaces of the pump. Edwards
recommend the use of non-CFC solvents such as isopropanol or ethanol. Only a minimal
amount of a cleaning soluon is required which is suitable for the contaminants on the
pump surfaces.
For environmental reasons, keep wastage of cleaning soluons and solvents to a
minimum.
8.4 Decoding service status words
The service status may be accessed directly via the serial link. This method of accessing
service status will give the most complete picture of current and future service
requirements and will allow preventave maintenance acvies to be scheduled.
A summary of the current pending service status is provided in response to the service
status command:
Command
? v 8 8 1 cr
The reply will be as follows:
08/2021 - ©Edwards Limited
Page 90B8J200880_D
300812671_002_C3
B8J200880_D - Maintenance