If the LEDs do not light up as expected, or if the red or yellow LEDs begins to ash a
repea
ted sequence or if the red LED is lit, refer to Fault nding on page 96. If the bearing
LED is lit or ashing refer to Bearing monitoring and condioning system on page 106.
7.2 Vent opons, vent valve connecon and control
Note:
If manually venng the pump when it is at full rotaonal speed and the sub
sequent rate
of pressure rise is too high, the pump life may be reduced. When using the manual vent
valve supplied, it is recommended to either limit the vent or only open the vent valve
aer the pump speed has fallen to 50% of full rotaonal speed. Do not vent from the
backing line as this may lead to contaminaon. If venng into the vacuum system and
using an oil sealed rotary backing pump, select a point upstream of the pump, to prevent
oil back-streaming from the backing line. Table: Vent restrictor orice diameter if venng
the vacuum system chamber gives an indicaon of the appropriate orice size to be
ed to the vent-valve for given vacuum system volumes in order that the vent rate is
kept within the limits given in Vent gas specicaon and vent control data on page 30.
To maintain the cleanliness of the vacuum system, venng the pump (or vacuum system)
is recommended whenever switching the pump o.
7.2.1 Manual vent valve
Edwards recommends that the manual vent valve is opened only aer the pump speed
has fallen to 50% of full rotaonal speed.
A manual vent valve is supplied with the nEXT pump. It is not possible to accurately
control the rate of pressure rise using the manual vent valve so take care not to open it
too quickly.
7.2.2 TAV5 or TAV6 solenoid vent valve
The TAV solenoid valves can be purchased as accessories and come either pre-wired
with an auxiliary connector ed or as a bare-wire opon.
Both 'Normally Closed' and 'Normally Open' valves are available, refer to Accessories.
Venng may be accomplished by one of the following means, and by using the
procedure below:
▪ Use the TAV solenoid vent valve in the manual vent valve posion.
▪ Use the TAV solenoid vent valve connected to a convenient ange on the v
acuum
system.
▪ Use an alternave vent valve, with an appropriate restricon, connected to the
vacuum system.
When using the TAV vent valve, the pump may only be hard vented when it is at full
speed if the vacuum system has a volume of 5 litres or more.
If the volume of the vacuum system is less than 5 litres, incorporate a suitable vent
restrictor and vent the pump when it is at full speed, or use the controlled venng
opon. Table: Vent restrictor orice diameter if venng the vacuum system chamber
gives an indicaon of the appropriate orice size to be ed to the vent valve for given
08/2021 - ©Edwards Limited
Page 76B8J200880_D
300812671_002_C3
B8J200880_D - Operaon