The reply will be as follows:
Command
* S 8 5 6 sp r cr
The normal speed is now stored in memory within the nEXT pump.
To check wha
t normal speed is set, send a query as follows:
Command
? S 8 5 6 cr
The reply will be as follows:
Reply
= S 8 5 6 sp d d d cr
6.1.8  Timer seng and opons
The meout period is a user-congurable opon and can be set to any value from
1 to 30 minutes.
When the pump is shipped, it is congured with a default meout period of 8 minutes.
Refer to Timer on page 16 for a full descripon of mer funconality.
To change the mer seng, send the following command (where the characters 'd'
represent the meout period in minutes):
Command
! S 8 5 4 sp d d cr
The reply will be as follows:
Reply
* S 8 5 4 sp r cr
The Timer seng is now s
tored in memory within the nEXT pump.
To check what me‑out period is set, send a query as follows:
Command
? S 8 5 4 cr
The reply will be as follows:
Reply
= S 8 5 4 sp d d cr
The mer is permanently enabled during r
amp‑up, however it is oponal whether to
have it enabled at other mes. When the pump is shipped, the Timer is enabled by
default.
To disable the mer, send the following serial command:
Command
! S 8 7 0 sp 0 cr
The reply will be as follows:
Reply
* S 8 7 0 sp r cr
The state of the mer opon is st
ored in memory within the nEXT pump.
To enable the mer again, send the following serial command:
Command
! S 8 7 0 sp 1 cr
The reply will be as follows:
Reply
* S 8 7 0 sp r cr
08/2021 - ©Edwards Limited
Page 71B8J200880_D
300812671_002_C3
B8J200880_D - Conguraon