Bit number Status ag Acve ag means
13 Bearing condioning
aut
omac enable
Set if Bearing condioning is automacally
enabled on bit 8 or 9 set. Typically enabled
by default
14 Bearing condioning
enabled
Set if Bearing condioning is presently ena-
bled, refer to Bearing run in mode on page
107.
15-31 Reserved -
8.5 Controller run me
The run hours and recommended service me of the controller on the nEXT pump can
be monitored.
Send the following query:
Command
? v 8 8 2 cr
The reply will be as follows, where the rst number is the hours run by the controller
and the second is the number of hours unl service is recommended:
Reply
= v 8 8 2 sp d d d d d d ; d d d d d d cr
Note that the number of hours unl the nex
t service is due is esmated by the controller
based on the operang condions of the pump and therefore may reduce at more or
less than 1 hour per hour. The dominant factor in extending controller life is reduced
pump temperature.
8.6 Pump run me
The run hours and recommended service me of the rotor in the nEXT pump can be
monitored.
Send the following query:
Command
? v 8 8 4 cr
The reply will be as follows, where the rs
t number is the start-stop cycles completed by
the pump and the second is the number of start-stop cycles unl service is
recommended:
Reply
= v 8 8 4 sp d d d d d d ; d d d d d d cr
8.7 Pump cycles
The number of start-stop cycles completed and the number remaining unl the next
service is due can be monitored.
Send the following query:
Command
? v 8 8 4 cr
The reply will be as follows, where the rst number is the start-stop cycles completed by
the pump and the second is the number of start-stop cycles unl service is
recommended:
08/2021 - ©Edwards Limited
Page 92B8J200880_D
300812671_002_C3
B8J200880_D - Maintenance