EasyManua.ls Logo

Edwards nEXT730D - Page 91

Edwards nEXT730D
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reply
= v 8 8 1 sp h h h h h h h h cr
The service status word is made up of 8 hexidecimal digits. To decode this word, convert
each digit int
o a 4-digit binary number in the same manner as the system status word,
which is described in Decoding system status words on page 102.
Each binary digit (bit) represents a ag that is either acve (state 1) or not acve (state
0). To help decode the service status word, each bit is numbered (starng with 0 for the
least signicant to 31 for the most signicant) as shown in Decoding system status words
on page 102. The meaning of each bit in the service status word is given in Flashing
service codes on page 100.
Table 24
Service ags
Bit number Status ag Acve ag means
0 Reserved -
1 Bearing service due Set when hours unl bearing service due = 0
or bearing replace required due to inacvity
per bit 9 below
2 Pump service due Set when hours unl pump service due = 0
or cycles unl pump service due = 0
3 Controller service due Set when hours unl controller service due =
0
4 Reserved -
5 Reserved -
6 Reserved -
7 Service due Service is due, or underway per bit 12 below.
Specic operaon required should be deter-
mined by checking the bits above, and 10
and 11 below
8 Bearing condioning re-
quired ag
Set on exceeding Bearing Run In Mode
(BRIM) inacve period, typically 1 year.
Cleared on full BRIM compleon, provided
no step forward, or these three condions:
1) TMP start while bit 8 or 9 set, but without
full BRIM enabled.
2) BRIM disable while bit 8 or 9 set, whether
or not BRIM underway.
3) TMP stop while BRIM underway.
9 Bearing replace re-
quired ag
Set on exceeding Bearing Replace (BR) inac-
ve period, typically 3 years. Cleared on BR
conrm.
10 Bearing condioning re-
quired
Per bit 8, but suppressed if bit 11 or 12 set
11 BR required Per bit 9, but suppressed if bit 12 set
12 Bearing condioning
running
Set if Bearing condioning is presently run-
ning, from start to Post Ramp (PR) com-
mence
08/2021 - ©Edwards Limited