Reply
= v 8 8 1 sp h h h h h h h h cr
The service status word is made up of 8 hexidecimal digits. To decode this word, convert
each digit int
o a 4-digit binary number in the same manner as the system status word,
which is described in Decoding system status words on page 102.
Each binary digit (bit) represents a ag that is either acve (state 1) or not acve (state
0). To help decode the service status word, each bit is numbered (starng with 0 for the
least signicant to 31 for the most signicant) as shown in Decoding system status words
on page 102. The meaning of each bit in the service status word is given in Flashing
service codes on page 100.
Table 24
Service ags
Bit number Status ag Acve ag means
0 Reserved -
1 Bearing service due Set when hours unl bearing service due = 0
or bearing replace required due to inacvity
per bit 9 below
2 Pump service due Set when hours unl pump service due = 0 
or cycles unl pump service due = 0
3 Controller service due Set when hours unl controller service due =
0
4 Reserved -
5 Reserved -
6 Reserved -
7 Service due Service is due, or underway per bit 12 below.
Specic operaon required should be deter-
mined by checking the bits above, and 10
and 11 below
8 Bearing condioning re-
quired ag
Set on exceeding Bearing Run In Mode
(BRIM) inacve period, typically 1 year.
Cleared on full BRIM compleon, provided
no step forward, or these three condions: 
1) TMP start while bit 8 or 9 set, but without
full BRIM enabled. 
2) BRIM disable while bit 8 or 9 set, whether
or not BRIM underway. 
3) TMP stop while BRIM underway.
9 Bearing replace re-
quired ag
Set on exceeding Bearing Replace (BR) inac-
ve period, typically 3 years. Cleared on BR
conrm.
10 Bearing condioning re-
quired
Per bit 8, but suppressed if bit 11 or 12 set
11 BR required Per bit 9, but suppressed if bit 12 set
12 Bearing condioning
running
Set if Bearing condioning is presently run-
ning, from start to Post Ramp (PR) com-
mence
08/2021 - ©Edwards Limited
Page 91B8J200880_D
300812671_002_C3
B8J200880_D - Maintenance