Fault Yellow service LED is ashing a repea
ted sequence
Cause A service is required.
Remedy See Flashing servic
e codes on page 100.
Fault The red alarm LED is on
Cause Normal during the soware upgrade process. At other mes indicates a problem
has been detected in the FLASH memory.
Remedy If during sowar
e upgrade then complete the upgrade process. Otherwise cycle
power and if this does not help then perform a soware upgrade.
Fault The red alarm LED is ashing
Cause A fail c
ondion has bec
ome acve.
Remedy Note the posion of the long ashes within the series of 6 ashes to work out the
error code.
Look up the ashing error code in Flashing error codes on page 100 and follow the
advice given.
Fault Blue bearing LED is ashing a repeated sequence
Cause A bearing service due to inacvity is requir
ed.
Remedy See Flashing inacvity service codes on page 101
Fault Blue bearing LED is lit
Cause Bearing run in mode (BRIM) is running.
Remedy Wait unl BRIM is nished. This may take 4.3 hours. We recommend not to cancel
BRIM.
Fault Pump is not accelerang t
o full speed
Cause Bearing run in mode (BRIM) is running.
Remedy Wait unl BRIM is nished. This may t
ake 4.3 hours. We recommend not to cancel
BRIM, refer to Bearing run in mode on page 107.
Fault Any other problems
Cause Fault condions other than the ones pr
eviously menoned.
Remedy Contact the supplier or us.
11.1 Command set error codes
Error codes that may be returned for serial control and monitoring.
Table 26
Command error codes
Returned error code
Meaning
0 No error
08/2021 - ©Edwards Limited