The RS485/CAN/RS232 slide switch must be in the RS232 posion if either the TIC or TA
G
is to be used to control the nEXT pump.
5.4.3 Connect the logic interface to control equipment
To operate the pump using your own control system, use suitable connector mang half
(not supplied) to connect the control equipment to the connector on the logic interface
cable.
5.4.4 Connect the electrical supply
CAUTION:
This product r
equires a separate power supply (not included). The power supply
should be adequately protected against a hazardous live condion (for example, in
case of a short circuit).
CAUTION:
Incorpor
ate a suitable Emergency Stop Switch in the electrical supply. Locate the
switch in an easily accessible posion and mark it as the emergency disconnecng
device for the pump. Failure to do so will result in not being able to switch o the
pump in an emergency.
CAUTION:
Incorpor
ate a suitable fuse or current liming device in the 48 V d.c. supply line to
the pump. If this is not done and a fault develops, the pump may develop a
hazardous surface temperature or present a re hazard.
Refer to Table: Logic interface technical data for suitable fuse rangs.
CAUTION:
Do not ex
ceed the maximum supply voltage. Excessive supply voltage will cause
permanent damage to the control electronics and may result in a mechanical hazard
in some failure condions.
The electrical supply provided for the nEXT pump must meet the requirements of
UL61010-1 and EN61010-1. Ensure that hazardous voltages as dened in UL61010-1 and
EN61010-1 cannot be present on the electrical interface to the nEXT pump.
The nEXT pump 0 V is not referenced to earth (ground). Ensure that there is only one
path between 0 V and earth. Mulple connecons between 0V and earth must be
avoided in order to avoid unexpected oset voltages on control and status signals and
possible problems with serial communicaons. If no other connecon is present
between 0 V and earth, the connecon should be made at the power supply. Be aware
that other electrical equipment connected to the system could introduce a connecon
between 0 V and earth, for example a personal computer or measuring equipment.
To operate the nEXT pump using the customer’s own electrical supply, use a suitable
connector mang half (not supplied) to connect the electrical supply to the connector
on the pump power supply cable.