6.1.4 Powering a fan from the motor controller
A fan c
an be powered from the nEXT motor controller on auxiliary output 2.
Send the following command:
Command
! S 8 6 4 sp 8 cr
The reply will be in the following format:
Reply
* S 8 6 4 sp r cr
The permanently enabled fan seng is now stored in memory within the nEXT pump.
When the pump is shipped, it is set up to run a vent valve. Send a query to nd out what
the seng is as follows:
Command
? S 8 6 4 cr
The reply will be in the following format:
Reply
* S 8 6 4 sp d cr
If the character 'd' is 8, then the fan is enabled. If it is anything else, congure the motor
controller to run the fan.
6.1.5 Controlled venng opons
If the motor controller is used to automacally control a vent valve, there are a number
of venng opons available.
Refer to Table: Vent valve opons for available venng opons and refer to Table: Valve
types for available valve types.
To set a vent opon on auxiliary output 1 send the following command, (where the
character 'd' refers to the opon number shown in Table: Vent valve opons):
Command
! S 8 5 3 sp d cr
The reply will be in the following format:
Reply
* S 8 5 3 sp r cr
The venng
opon is now stored in memory within the nEXT pump.
To check what venng opon is set, send a query as follows:
Command
? S 8 5 3 cr
The reply will be in the following format:
Reply
= S 8 5 3 sp d cr
To set the valve type on auxiliary output 1, send the following command, (where the
charact
er 'd' refers to the opon number shown in Table: Valve types):
Reply
! S 8 7 7 sp d cr
The reply will be in the following format:
08/2021 - ©Edwards Limited
Page 69B8J200880_D
300812671_002_C3
B8J200880_D - Congur
aon