Cause The inlet pressure is poor.
Remedy
Remove the pump from the system and test the ulmat
e pressure of the
pump alone (refer to ).
If inlet pressure is poor, check the pump for contaminaon and refer to .
Leak test the pump. If the leak rate > 1x10
-7
mbar l s
-1
(1x10
-5
Pa l s
-1
),
contact the supplier or Edwards.
Fault The pump is very noisy or there is excessive vibraon or both
Cause The pump
rotaonal speed is the same as the r
esonant frequency of the aached
system.
Remedy Change the natural frequency of the system or isolate the pump using exible
bellows.
Cause The vibraon is being tr
ansmied from the rotary backing pump.
Remedy Fit exible bellows or a vibraon isolator in the backing line.
Cause The noise is irregular and geng pr
ogressively worse.
Remedy There is a defecve bearing. Contact the supplier or us.
Cause The pump is making a constant high pitched noise.
Remedy The rotor is out of balance. Contact the supplier or us.
Fault No serial comms
Cause No electrical supply, loss of serial link.
Remedy Check that the electrical supply is on, that the serial link is connected and that the
serial enable line is acve.
Cause Slide s
witch is not in the correct posion for R
S232 or RS485.
Remedy
Check that the electrical supply is on, that the serial link is connected and
that the serial enable line is acve.
Le for RS485, right for RS232.
Right for parallel control and monitoring.
Cause Incorrect baud rate and node address.
Remedy Check baud rate and, if operang in mul-dr
op mode, the node address matches
those of the pump.
Cause Check that the red light is not on or ashing.
Remedy If LED is on, perform the checks against the red alarm LED symptoms
Fault Fail signal or standby signal not working
Cause The serial enable line is inacve.
R
emedy Must be disconnected or driven high for use of standby and fail lines.
Cause Slide switch is not in the correct posion for R
S232 or RS485.
Remedy Le for RS485, right for RS232.
Right for parallel control and monitoring.
08/2021 - ©Edwards Limited
Page 98B8J200880_D
300812671_002_C3
B8J200880_D - Fault nding

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards nEXT730D and is the answer not in the manual?

Edwards nEXT730D Specifications

General IconGeneral
BrandEdwards
ModelnEXT730D
CategoryWater Pump
LanguageEnglish

Related product manuals