9. Storage
Av
oid long-term storage if possible. When long-term storage is necessary, the pump
should be set up and run for at least eight hours every 12 months.
Use the following procedure to store the pump.
1. Place protecve covers over all ports, the main inlet, inter-stage exhaust, purge
and vent ports.
2. Place the pump in its packing materials. For fastest pump-down when the pump is
put back into service, seal the pump inside a plasc bag together with a suitable
desiccant.
3. Store the pump in cool, dry condions, preferably not exposed to atmospheric air
unl required for use. When required, prepare and install the pump as described in
Installaon on page 50.