Reply
= S 8 5 0 sp 0 cr
If the pump already has a mul-drop addr
ess, there will be no reply. Communicate with
the pump in mul-drop message protocol. Refer to Mul-drop operaon on page 46 for
more informaon about mul-drop mode and mul-drop message protocol.
Use the following query (using wild card address 99 which means 'any' node) to nd out
the mul-drop address of the nEXT pump:
Command
# 9 9 : 9 9 ? S 8 5 0 cr
The reply will be as follows, where dd denotes the mul-drop address of the pump:
Reply
# 9 9 : 9 9 = S 8 5 0 sp d d cr
Mul-drop mode can be disabled by assigning the pump an address 0. To do this, send
the following command (where dd denotes the mul-drop address of the pump and XX
denotes the address of the node that is sending the command):
Command
# d d : x x 1 S 8 5 0 sp 0 cr
The reply will be as follows:
Reply
# x x : d d * S 8 5 0 sp 0 cr
Once mul-drop mode is disabled, the pump will no longer respond to mul-drop
commands.
6.1.13 Congure the pump using a TIC
The nEXT pump can be congured using the Edwards TIC Turbo and Instrument
Controller or TIC Turbo Controller.
Refer to TIC or TAG logic interface connecons on page 56.
It is possible to set the following parameters of the nEXT pump using the TIC:
▪ Power limit seng
▪ Vent opons and ven
t type, including running a fan from the motor controller
▪ Standby speed seng
▪ Normal speed seng
▪ Timer sengs - both enabling/disabling the mer AND
seng the me‑out period
▪ Electronic braking opons
▪ Factory default sengs
For informaon on how to perform these sengs, refer to the TIC Turbo and Instrument
Controller or TIC Turbo Controller Instrucon Manuals.
Note:
It is not possible to c
ongure the Analogue Output Opons using the TIC. There is also no
facility within the TIC to assign a mul‑drop address to the nEXT pump.