Edwards nEXT730D - Remove and Replace the Inlet Screen

Edwards nEXT730D
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAUTION
Ensure tha
t any cabling and or pipe work aached to the pump are routed carefully
to avoid causing a slip/trip hazard and to prevent any damage to the cable.
Figure 16 Connecons and controls
67
89
1011
12
1
2
3
4
5
1. Inlet ange 2. 4x Boreholes for transport eyes M8, 14
deep
3. Purge port G 1/8", closed with plug
screw
4. Backing port
5. Grounding connecon
6. Bearing s
tatus indicaon (indicates when
bearing service is recommended)
7. Control interface
8. Accessory connector
9. 48 V d.c. power supply 10. LED indicators on drive unit
11. Vent port G 1/8", closed with plug screw 12. RS485/CAN/RS232 slide switch, USB
connector, Standby speed change
buons
1. Inlet ange 2. 4x Boreholes for transport eyes M8, 14
deep
3. Purge port G 1/8", closed with plug
screw
4. Backing port
5. Grounding connecon
6. Bearing status indicaon (indicates when
bearing service is recommended)
7. Control interface
8. Accessory connector
9. 48 V d.c. power supply 10. LED indicators on drive unit
11. Vent port G 1/8", closed with plug screw 12. RS485/CAN/RS232 slide switch, USB
connector, Standby speed change
buons
5.2.1 Remove and replace the inlet screen
The inlet screen is supplied ed on CF pumps only. Remove the inlet screen only if
there is no possibility that debris can fall into the pump. If the inlet screen is removed,
the pumping speed will increase by up to 20%.
CAUTION
The rot
or blades on the pump are very sharp. The absence or removal of inlet screens
or strainers exposes the risk of injury from sharp edges or moving parts.
Care must be taken to avoid foreign objects from entering the pump during inst
allaon.
The pump is supplied with either an integral mesh centring O-ring seal or a pre-ed
mesh screen. This must be ed during installaon.
In the unlikely event of pump fracture, it is possible that the inlet screen may not trap all
of the debris within the pump. Ensure that the system can contain any debris that may
escape from the pump.
The screen protects the pump from contaminaon. Do not remove the inlet screen unl
the pump is about to be mounted onto the system.
1. To remove the inlet screen from a pump with CF inlet ange, use a ben
t wire hook
or small screwdriver to carefully lever the inlet screen out from the inlet ange.
08/2021 - ©Edwards Limited
Page 52B8J200880_D
300812671_002_C3
B8J200880_D - Installaon