EasyManuals Logo

Edwards nEXT730D User Manual

Edwards nEXT730D
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
nEXT
730D 730D 930D 930D
Power consumpon a
t ulmate pres-
sure
W 40
Type of protecon
IP
Type 1 (UL
50E)
54
1
Cooling standard Convecon*
Cooling oponal Air or water*
Cooling water c
onnecon
plug-in connecon f
or 6x1 hose / alternave G
1/8"
Cooling water consumpon l/h 60
P
ermissible cooling water pressure
(bar(g): bar (
gauge) is overpressure,
i.e. atmospheric pressure = 0 bar(g))
bar(g) 6
Permissible cooling water tempera-
ture
°C 15 to 35
Weight kg 14.6 19.6 15.4 21.7
Recommended forevacuum pumps XDS35i or E2M28
Noise level
with c
on
vecon cooling
with radial air cooler
dB(A) < 40
< 55
Max. bake-out temperature of the CF
version, -water cooled
°C 100
Purge gas ow
mbar · I · s
-1
sccm
0.4
24
Pur
ge gas connecon G 1/8"
V
enng connecon G 1/8"
* Depending on the ambien
t temperature, the gas throughput and the type of gas, forced
air c
ooling or wat
er cooling may be necessary.
Table 4
Technical data nEXT1230H
nEXT
1230H
High-vacuum connecon DN 200 CF 200 ISO-F 200 ISO-K
F
orevacuum connecon DN 40 KF
Pumping speed f
or
N
2
Ar
He
H
2
I · s
-1
1250
1150
1350
1150
08/2021 - ©Edwards Limited
Page 23B8J200880_D
300812671_002_C3
B8J200880_D - Technical data