Connect the cooling water hoses.
The hose connecons ma
y be unscrewed and removed to make use of the integrated G
1/8" threads.
Ensure that the water supply is constant and within the quality, temperature and ow
rate limits specied in Table: Water cooling block supply requirements on page 31.
NOTICE:
In the ev
ent of coolant failure, the pump must be allowed to cool down to ambient
temperature before restarngs.
NOTICE:
Condensaon may occur if a w
ater cooling block is used in high humidity
environments. Protect the pump and other equipment by considering the design of
the installaon.
Pipes in the water cooling circuit may become blocked if the cooling water contains too
much calcium carbonate or if it contains parcles that are too large. Corrosion of the
water cooling circuit may occur if there is too lile calcium carbonate and oxygen in the
water. Good quality drinking water is usually suitable. If in doubt, check the quality of
the cooling water supply and, if necessary, provide treatment and ltraon.
Either of the two push t connectors on the water cooler can be used for the water
supply or return connecons.