PROline Prosonic Flow 93 FOUNDATION Fieldbus 3 Transducer Blocks (transmission blocks)
Endress+Hauser 113
System Param. -
Positive Zero
Return
AUTO - OOS This parameter can be used to interrupt evaluation of measured
variables. This is necessary when a piping system is being
cleaned, for example. The setting acts on all parameters and
calculations of the measuring device.
Options:
OFF
(signal output not interrupted)
ON
(signal output is set to the value 0)
Factory setting:
OFF
Note!
When positive zero return is active, a flow value of "0" is output
via the output value OUT (AI Block).
Active positive zero return is relayed to downstream function
blocks or higher-level process control systems by means of the
status UNCERTAIN for the output value OUT (AI Block).
Positive zero return can be also be controlled using cyclic data
transmission via the Discrete Output function block, (see DO
function block Page 181).
Adjustment -
Adjust Zeropoint
AUTO - OOS This parameter can be used to start zero point adjustment
automatically. The new zero point determined by the measuring
system is adopted by the Calibration Data - Zero Point
parameter, (see Page 126).
Options:
CANCEL
START
Factory setting:
CANCEL
Caution!
Before carrying this out, please refer to the Operating Instructions
for PROline Prosonic Flow 93 FOUNDATION Fieldbus, (BA078D/
06/en), for a detailed description of the procedure for zero point
adjustment.
TRANSDUCER_CH1 and TRANSDUCER_CH2 (E+H parameters) /
Base index 1200 and 1300
Parameter Write access
with operating
mode
(MODE_BLK)
Description