Commissioning Proline Prosonic Flow W 400 HART
106 Endress+Hauser
Totalizer 1 to n operation mode
→  106
Totalizer 1 to n failure behavior
→  106
Parameter overview with brief description
Parameter Prerequisite Description Selection Factory setting
Assign process variable 1 to n Select process variable for
totalizer.
Off
Volume flow
Mass flow
Volume flow
Process variable unit 1 to n A process variable is selected
in the Assign process variable
parameter (→  106) of the
Totalizer 1 to n submenu.
Select the unit for the process
variable of the totalizer.
g
*
kg
*
t
*
oz
*
lb
*
STon
*
cm³
*
dm³
*
*
ml
*
l
*
hl
*
Ml Mega
*
af
*
ft³
*
Mft³
*
Mft³
*
fl oz (us)
*
gal (us)
*
kgal (us)
*
Mgal (us)
*
bbl (us;liq.)
*
bbl (us;beer)
*
bbl (us;oil)
*
bbl (us;tank)
*
gal (imp)
*
Mgal (imp)
*
bbl (imp;beer)
*
bbl (imp;oil)
*
None
*
Depends on country:
ft³
Totalizer 1 to n operation mode A process variable is selected
in the Assign process variable
parameter (→  106) of the
Totalizer 1 to n submenu.
Select totalizer operation
mode, e.g. only totalize
forward flow or only totalize
reverse flow.
Net
Forward
Reverse
Net flow total
Totalizer 1 to n failure behavior A process variable is selected
in the Assign process variable
parameter (→  106) of the
Totalizer 1 to n submenu.
Select totalizer behavior in the
event of a device alarm.
Hold
Continue
Last valid value +
continue
Stop
* Visibility depends on order options or device settings