Proline Prosonic Flow W 400 HART Commissioning
Endress+Hauser 115
Diagnostic event category
→  115
Diagnostic event simulation
→  115
Parameter overview with brief description
Parameter Prerequisite Description Selection / User entry
Assign simulation process variable Select a process variable for the
simulation process that is activated.
Off
Volume flow
Mass flow
Flow velocity
Sound velocity
Temperature
Density
Process variable value A process variable is selected in the
Assign simulation process variable
parameter (→  115).
Enter the simulation value for the
selected process variable.
Depends on the process
variable selected
Status input 1 simulation For the following order code:
"Output; input", option I "4-20mA HART,
2x pul./freq./switch output; status
input"
Switch simulation of the status input on
and off.
Off
On
Input signal level 1 In the Status input simulation
parameter, the On option is selected.
Select the signal level for the simulation
of the status input.
High
Low
Current output 1 simulation Switch the simulation of the current
output on and off.
Off
On
Current output value In the Current output simulation
parameter, the On option is selected.
Enter the current value for simulation. 3.59 to 22.5 mA
Frequency output 1 to n simulation In the Operating mode parameter, the
Frequency option is selected.
Switch the simulation of the frequency
output on and off.
Off
On
Frequency output 1 to n value In the Frequency simulation 1 to n
parameter, the On option is selected.
Enter the frequency value for the
simulation.
0.0 to 12 500.0 Hz
Pulse output simulation 1 to n In the Operating mode parameter, the
Pulse option is selected.
Set and switch off the pulse output
simulation.
For Fixed value option: Pulse
width parameter (→  96)
defines the pulse width of the
pulses output.
Off
Fixed value
Down-counting value
Pulse value 1 to n In the Pulse output simulation 1 to n
parameter, the Down-counting value
option is selected.
Enter the number of pulses for
simulation.
0 to 65 535
Switch output simulation 1 to n In the Operating mode parameter, the
Switch option is selected.
Switch the simulation of the switch
output on and off.
Off
On
Switch state 1 to n Select the status of the status output for
the simulation.
Open
Closed
Device alarm simulation Switch the device alarm on and off. Off
On
Diagnostic event category Select a diagnostic event category. Sensor
Electronics
Configuration
Process
Diagnostic event simulation Select a diagnostic event to simulate
this event.
Off
Diagnostic event picklist
(depends on the category
selected)