Mounting procedure Proline Prosonic Flow W 400 HART
32 Endress+Hauser
A0043373
19 Holder with strapping bands (medium nominal diameters), with hinged screw
1 Mounting bolts
2 Strapping band
3 Tensioning screw
A0044350
20 Holder with strapping bands (medium nominal diameters), without hinged screw
1 Mounting bolts
2 Strapping band
3 Tensioning screw
Sensor holder with strapping bands (large nominal diameters) )
Can be used for
• Measuring devices with measuring range DN 50 to 4000 (2 to 160")
• Mounting on pipes DN > 600 (24")
• 1-traverse mounting or 2-traverse mounting with 180° arrangement
• 2-traverse mounting with two-path measurement and 90° arrangement (instead of
180°)
A0044648
Procedure:
1. Measure the pipe circumference. Note down the full/half or quarter circumference.
2. Shorten the strapping bands to the required length (= measuring pipe circumference
+ 30 mm (1.18 in)) and trim the cut edges.