EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow W 400 - Page 32

Endress+Hauser Proline Prosonic Flow W 400
178 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mounting procedure Proline Prosonic Flow W 400 HART
32 Endress+Hauser
1
2
3
A0043373
 19 Holder with strapping bands (medium nominal diameters), with hinged screw
1 Mounting bolts
2 Strapping band
3 Tensioning screw
1
2 3
A0044350
 20 Holder with strapping bands (medium nominal diameters), without hinged screw
1 Mounting bolts
2 Strapping band
3 Tensioning screw
Sensor holder with strapping bands (large nominal diameters) )
Can be used for
Measuring devices with measuring range DN 50 to 4000 (2 to 160")
Mounting on pipes DN > 600 (24")
1-traverse mounting or 2-traverse mounting with 180° arrangement
2-traverse mounting with two-path measurement and 90° arrangement (instead of
180°)
A0044648
Procedure:
1. Measure the pipe circumference. Note down the full/half or quarter circumference.
2. Shorten the strapping bands to the required length (= measuring pipe circumference
+ 30 mm (1.18 in)) and trim the cut edges.