System integration Proline Prosonic Flow W 400 HART
82 Endress+Hauser
Parameter Description Selection / User entry
Burst variable 0 For HART command 9 and 33: select the HART device variable
or the process variable.
Not used
Volume flow
Mass flow
Flow velocity
Sound velocity
Temperature
Density
Signal strength
*
Signal to noise ratio
*
Acceptance rate
*
Turbulence
*
Totalizer 1
Totalizer 2
Totalizer 3
Percent of range
Measured current
Primary variable (PV)
Secondary variable (SV)
Tertiary variable (TV)
Quaternary variable (QV)
Burst variable 1 For HART command 9 and 33: select the HART device variable
or the process variable.
See the Burst variable 0 parameter.
Burst variable 2 For HART command 9 and 33: select the HART device variable
or the process variable.
See the Burst variable 0 parameter.
Burst variable 3 For HART command 9 and 33: select the HART device variable
or the process variable.
See the Burst variable 0 parameter.
Burst variable 4 For HART command 9: select the HART device variable or the
process variable.
See the Burst variable 0 parameter.
Burst variable 5 For HART command 9: select the HART device variable or the
process variable.
See the Burst variable 0 parameter.
Burst variable 6 For HART command 9: select the HART device variable or the
process variable.
See the Burst variable 0 parameter.
Burst variable 7 For HART command 9: select the HART device variable or the
process variable.
See the Burst variable 0 parameter.
Burst trigger mode Select the event that triggers burst message X. Continuous
Window
*
Rising
*
Falling
*
On change
Burst trigger level Enter the burst trigger value.
Together with the option selected in the Burst trigger mode
parameter the burst trigger value determines the time of burst
message X.
Signed floating-point number
Min. update period Enter the minimum time span between two burst commands of
burst message X.
Positive integer
Max. update period Enter the maximum time span between two burst commands of
burst message X.
Positive integer
* Visibility depends on order options or device settings