Proline Prowirl R 200 HART Commissioning
Endress+Hauser 107
Parameter Prerequisite Description Selection / User
entry
Factory setting
Mol% i-C4H10 The following conditions are
met:
• In the Select medium
parameter, the Gas option is
selected.
• In the Select gas type
parameter, the Natural gas
option is selected.
• In the Density calculation
parameter, the ISO 12213-
2 option is selected.
Enter amount of substance for
the gas mixture.
0 to 100 % 0 %
Mol% i-C5H12 The following conditions are
met:
• In the Select medium
parameter, the Gas option is
selected.
• In the Select gas type
parameter, the Natural gas
option is selected.
• In the Density calculation
parameter, the ISO 12213-
2 option is selected.
Enter amount of substance for
the gas mixture.
0 to 100 % 0 %
Mol% Kr The following conditions are
met:
• In the Select medium
parameter, the Gas option is
selected.
• In the Select gas type
parameter, the Gas mixture
option is selected.
• In the Gas mixture
parameter, the Krypton Kr
option is selected.
Enter amount of substance for
the gas mixture.
0 to 100 % 0 %
Mol% N2 The following conditions are
met:
In the Select medium
parameter, the Gas option is
selected.
– In the Select gas type
parameter, the Gas mixture
option is selected and in the
Gas mixture parameter, the
Nitrogen N2 option is
selected.
Or
– In the Select gas type
parameter, the Natural gas
option is selected and in the
Density calculation
parameter, the AGA Nx19
option or the ISO 12213- 2
option is selected.
Enter amount of substance for
the gas mixture.
0 to 100 % 0 %
Mol% n-C10H22 The following conditions are
met:
• In the Select medium
parameter, the Gas option is
selected.
• In the Select gas type
parameter, the Natural gas
option is selected.
• In the Density calculation
parameter, the ISO 12213-
2 option is selected.
Enter amount of substance for
the gas mixture.
0 to 100 % 0 %