Proline Prowirl R 200 HART Commissioning
Endress+Hauser 133
Energy flow
Medium Fluid Standards Explanation Heat/energy option
Steam
1)
IAPWS-
IF97/ASME
For fixed process
pressure or if the
pressure is read in via
current input/HART
Heat
Gross calorific value
2)
in relation to mass
Net calorific value
3)
in relation to mass
Gross calorific value
2)
in relation to corrected
volume
Net calorific value
3)
in relation to corrected
volume
Gas
Single gas
ISO 6976
Contains GPA 2172
For fixed process
pressure or if the
pressure is read in
via current input/
HART
Gas
mixture
ISO 6976 Contains GPA 2172
For fixed process
pressure or if the
pressure is read in
via current input/
HART
Air NEL40 For fixed process
pressure or if the
pressure is read in via
current input/HART
Natural
gas
ISO 6976 Contains GPA 2172
For fixed process
pressure or if the
pressure is read in
via current input/
HART
AGA 5
Liquids
Water IAPWS-
IF97/ASME
Liquefied
gas
ISO 6976 Contains GPA 2172
Other
liquid
Linear
equation
1) The measuring device is capable of calculating the volume flow, and other measured variables derived from
the volume flow, across all steam types with full compensation using the pressure and temperature. To
configure device behavior →  110
2) Gross calorific value: combustion energy + condensation energy of the flue gas (gross calorific value > net
calorific value)
3) Net calorific value: only combustion energy
Mass flow and energy flow calculation
NOTICE
The process pressure (p) in the process pipe is required to calculate the process
variables and the limit values of the measuring range.
With the HART device, the process pressure can be entered via the 4 to 20mA current
input or via HART from an external pressure measuring device (e. g. Cerabar M) or
entered as a fixed value in the External compensation submenu (→  110).