Test Setup - Test Configurator, Timer, and System
206 88000 Series
External Reference
External Reference
Allows the configuration of the external reference signal. Available with
Wander test application.
From the Test menu, tap Setup, Test Configurator, and tap on the
External Reference block.
Reference
Mode is set to External Reference.
Ext Reference In
Allow the selection of the reference signal for TE/TIE measurements.
Signal allows the selection of the reference interface: 2MHz, 10 MHz,
and 1PPS. Only 1PPS reference is available for 1PPS interface/rate test.
Ext Ref In on a green background indicates that a valid reference is
received.
LOS on a red background indicates that the received signal is not valid.
Connector indicates the connector used for the external reference
signal: RJ48C or BNC.
Frequency (MHz) displays the frequency of the received reference
signal.