Power Failure Recovery
744 88000 Series
Enabling Power Failure Recovery
Enabling Power Failure Recovery
To enable the automatic power failure recovery:
1. Enable launching the application when starting the platform (refer to
the platform user guide for more information):
From ToolBox X, tap on the System Settings button, the Startup
Applications button, and select the corresponding module’s check
box.
2. moduleEnable the platform automatic power on feature (refer to the
platform user guide for more information):
2a. From ToolBox X, tap on the System Settings button and the
ToolBox X button.
2b. Select the Power on the unit when AC outlet is connected or
after power outage check box.
3. Make sure that Windows does not require a user name and password.
The platform is set to require user name and password by default. To
disable Windows user name and password:
3a. From ToolBox X, tap on the System Settings button and the
Automatic Logon button.
3b. Clear the User must enter a user name and password to use
this computer check box and enter the password to confirm.
Note: The power failure recovery is not used when the application closes
normally.