Test Setup - Test Applications
Power Blazer 87
Wander
Wander
The Wander application provides data acquisition and processing
capabilities to quantify and validate the Time Error / Time Interval Error of a
synchronization equipment.
Time Error is defined as the difference between the time reported by a
clock signal under test and the time reported by a reference clock.
Typical Wander test applications:
Supported Interfaces/Rates: 100/1000M Electrical, 10GE LAN,
1GE Optical, 100M Optical, DS1, E1, 2 MHz, 10 MHz, and 1PPS.
Note: Only one instance of the Wander application is allowed on a platform due
to high processing activity related to MTIE calculation.
Note: The Wander test automatically stops when one of the following alarms
occur: LOS, OOF (DS1), LOF (E1), AIS (DS1/E1), or REF-FAULT.