264
Cause of the fault / solution
or switch on the
temperature sensor.
The measured
temperature is beyond
the technical limit of the
temperature sensor.
Temperature measurement range: -15 °C to +550 °C
No memory left
for saving data
Cannot create any more
measurement, balancing
step or new data block
for run up/coast down
because memory is full.
Transfer the data to Trendline and transfer the last balancing
configuration or route back to the Detector.
The battery is nearly
empty.
Please recharge the accumulator before continuing to
work with the Detector. Otherwise, the Detector could shut
down without any warning so as to protect the accumulator.
The battery is
empty Please
charge
The battery is empty,
the Detector cannot be
switched on.
The battery must be recharged .
No memory has been
allocated for
configurations that have
been sent to the
Detector.
The Trendline database may contain an error. Please
contact support@fis-services.com.
No balancing, CM or run
up/coast down template
has been saved.
You are trying to perform a free measurement, but there is
no template is stored in the Detector. Templates are only
attached if you click in Trendline on Detector > Send route
.
No balancing, CM or run
up/coast down
configuration has been
saved.
You wish to carry out balancing, a CM measurement or a
run up/coast down procedure but have not yet transferred a
configuration to the Detector. In Trendline click on Detector
> Send configuration or Detector > Send route.
No free measurements
have been performed
yet.
You are attempting to view free balancing, CM
measurements or run up/coast down measurements on the
Detector although neither of these types of measurement has
been carried out.
No balancing
configuration
stored.
The Detector is
attempting to save time
signals but cannot
because of insufficient
memory.
Send a balancing configuration to the Detector (click in
Trendline on Detector > Send configuration or
Detector > Send route) before you select a configuration
from the balancing menu of the Detector.
No memory
available for time
The Detector is
attempting to save time
This can occur if you want to save only those time signals
occurring in the event of an alarm. It is also possible that
218