265
Detector III
Error message
Description
Cause of the fault / solution
signals
signals but cannot as
there is not enough
memory.
you have taken multiple measurements at this measuring
point. The characteristic values are always saved, but none
or only some of the time signals are saved.
Input
overmodulated!
The vibration signal is
too large to be measured
by the Detector.
There are two possible causes for this:
The input signal is very large (>50 g). Use another
sensor (with a sensitivity of 10 mV/g, for example).
The strength of the signal increased after the amplifier
was adjusted due to an impact on the machine, for
example. The vibrations must remain more or less
constant throughout the entire measurement.
No RPM signal!
The rotational speed
signal is missing.
Possible causes:
The trigger sensor is not correctly aligned with the reflex
mark.
The reflex mark was not glued on.
The trigger sensor is too close to the reflex mark (<10
cm). Increase the distance.
The trigger sensor has been incorrectly configured
in Trendline (incorrect supply voltage, for example).
You have not connected the trigger sensor to the AUX
port of the Detector.
Rotational speed
unstable!
The rotational speed
signal is fluctuating.
Possible causes:
The trigger sensor is not correctly aligned with the reflex
mark.
The trigger sensor is too close to the reflex mark (should
be at least 10 cm).
Machine rotational speed fluctuates.
The rotational
speed not within
range.
The acquired speed is
outside the rotational
speed band defined in .
Reduce the speed of the machine or use the current
rotational speed as the new nominal rotational speed.
No conf. for RFID
No configuration found
for the RFID entry.
This error occurs when Trendline sends the RFID status for
a measuring point before a balancing, CM or run up/coast
down configuration was sent. Make sure that there is a
configuration for the measuring point on the Detector. The
Trendline database may contain an error. Please contact
support@fis-services.com.
Could not find at
least one
configuration
Detected RFID tags and
configured measuring
points do not match.
The Detector found at least one RFID tag without a
measuring point configuration. The missing measuring point
is therefore not offered for selection for the measurement.
Multiple IDs read.
Please repeat
Two or more tags were
found while trying to
assign an RFID tag.
The assignment is not possible as the Detector found at least
two tags in the RFID Reader read area. Remove the
unnecessary RFID tags from the read area.