266
Error message
Description
Cause of the fault / solution
ID already in use.
Please repeat
An RFID tag was read
that is already assigned
to another measuring
point.
Assign a new RFID tag to the current measuring point.
Change sensor to:
The next measurement
must be performed with
a different sensor.
If you have configured sensors with different sensitivities for
measuring points, you may need to change the sensor
between two measurements. In this case, the Detector tells
you which sensor you need to connect for the next
measurement.
Maximum number
of data reached!
Maximum number of
data points reached (in
run up/coast down).
The system aborts after 65535 values in a run up/coast
down. The values measured so far are saved.
Resonant
frequency!
Machine in resonant
range for balancing
measurement.
If you have defined one or more resonant ranges for a
machine in the Trendline software (see "setting up run up/
coast down " and "determining the resonant range of a
machine "), the Detector displays this message during
balancing if the current rotational speed is within such a
resonant range. You can continue the balancing
measurement. However, observe the information given in
balancing measurement .
Rotational speed
to high (low) for
Autostart /
Autostop
Rotational speed already
exceeds the selected
autostart and/or autostop
speed at run up.
Or: Rotational speed is
already lower than the
selected autostart and/or
autostop speed at coast
down.
Run up: If rotational speed is too high for an automatic start
and/or automatic stop, the system stops the automatic
measurement start and/or stop. You can start and/or stop
manually by pressing Enter.
Coast down: If rotational speed is too low for an automatic
start and/or automatic stop, the system stops the automatic
measurement start and/or stop. You can start and/or stop
manually by pressing Enter.
HW0 to HW2
(backup battery)
Self-test error
At start-up, the Detector performs a self-test to check internal
voltages. If one of these three messages is displayed, you
must replace the backup battery. Please contact
support@fis-services.com and tell us the error number. We
will then give you further details on how to replace the
battery. You can bypass the error message by pressing
Enter.
HW3 to HW7
(Internal error)
Self-test error
This is a critical error. Please contact support@fis-services.
com and tell us the error number. We will then give you
further information.
82
247
234