COMMAND DETAILS
35
Return Parameter <NR2> I unit = I volt Von value (volts)
1 1 volts
Query Example :CONF:VOLT:ON?
0.03
Von is set as 30 millivolts
(0.03 volts).
:CONFigure:VOLTage:RANGe
Channel Specific
Command
Description
Sets Voltage range for CC mode.
Syntax :CONFigure:VOLTage:RANGe <NRf>[V], L, H
Parameter <NRf>[ V] , L, H Range
16 Low range*
80V High range*
L Low range
H High range
*Load module dependent, PEL-2020 shown.
Example :CONF:VOLT:RANG L
Sets the range to Low for
the channel.
Query Syntax :CONFigure:VOLTage:RANGe?
Return Parameter <NR2> Range
16 Low PEL-2020,2030,2040
125 Low PEL-2041
80 High PEL-2020,2030,2040
500 High PEL-2041
Query Example :CONF:VOLT:RANG?
500
Returns the voltage
range. In this case high
for the PEL-2041.
:CONFigure:VOLTage:LATch
Channel Specific
Command
Description
Turn Von Latch on or off for the specific channel.
Syntax :CONFigure:VOLTage:LATch OFF/0, ON/1
PEL-2000 Programming Manual
36
Parameter
OFF/0, ON/1 Von Latch
OFF/0 Off
ON/1 On
Example :CONF:VOLT:LAT 1
Sets Von latch to ON.
Query Syntax :CONFigure:VOLTage:LATch?
Return Parameter <NR1> Von latch status
0 Latched Off
1 Latched On
Query Example :CONF:VOLT:LAT?
1
Von latch is set to ON.
:CONFigure:AUTO:LOAD
All channels
Description
Configures the load generator for Auto Load On or
Off at start up.
Syntax :CONFigure:AUTO:LOAD OFF/0, ON/1
Parameter OFF/0, ON/1 Auto Load
OFF/0 Off
ON/1 On
Example :CONF:AUTO:LOAD ON
Configures Auto Load to
On
Query Syntax :CONFigure:AUTO:LOAD?
Return Parameter <NR1> Auto Load Status
0 Off
1 On
Query Example :CONF:AUTO:LOAD?
1
Auto load is On.