COMMAND DETAILS
95
Query Example :RES:DYN:LOW:T1?
2.5
:CURR:DYN:LOW:T1?
MIN
0.000025
Returns the T1 time of
2.5 seconds.
Returns the minimum T1
time allowable for the
specific channel.
:RESistance:DYNamic:HIGH:L1/L2
Channel Specific
Command
Description
Sets the high range resistance levels (Level 1 & 2)
for CR dynamic mode.
Syntax :RESistance:DYNamic:HIGH:L1/L2 <NRf+>[OHM]
Parameter NRf+[OHM] Resistance
L1 10 Sets L1 to 10 ohms. (high
range only)
L2 20OHM Sets L2 to 20 ohms. (high
range only)
L1/L2 MIN Sets L1 or L2 to the
minimum level for the
specific channel.
L1/L2 MAX Sets L1 or L2 to the
maximum level for the
specific channel.
Example :RES:DYN:HIGH:L1 10
In high range CR
dynamic mode, Set L1
(level 1) to 10 ohms.
Query Syntax :RESistance:DYNamic:HIGH:L1/L2? MIN, MAX
Return Parameter <NR2>[MAX, MIN] Resistance
1 unit= 1 ohm Returns the resistance of
Level 1/ 2 (L1/L2).
Query Example :RES:DYN:HIGH:L2? MAX
15000.0
Returns the maximum
resistance allowed for
the channel. (PEL-2020)
PEL-2000 Programming Manual
96
:RESistance:DYNamic:HIGH:RISE/FALL
Channel Specific
Command
Description
Sets the high range rising/falling slew rate for CR
dynamic mode for the specific channel.
Syntax :RESistance:DYNamic:HIGH:RISE/FALL
<NRf+>[A/uS]
Parameter <NRf+>[A/uS] Slew Rate
RISE/FALL 0.8A/uS Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL 1 Sets the rising/falling slew
rate to 1A/uS
RISE/FALL MIN Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX Sets to the fastest
rising/falling slew rate.
Example :RES:DYN:HIGH:RISE 1.1
Sets the rising slew rate
to 1.1A/uS.
Query Syntax :RESistance:DYNamic:HIGH:FALL? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] Slew rate
1 Unit=1 amp/uS Returns the slew rate.
MAX/MIN Returns the allowable
maximum and minimum.
Query Example :RES:DYN:HIGH:FALL?
MAX
0.8000
For high range dynamic
CR mode, the minimum
value for the falling slew
rate is 0.8 A/uS for the
specific channel.