COMMAND DETAILS
37
:CONFigure:AUTO:MODE
All channels
Description
Configures the Auto Load mode as (run) Program
or Load.
Syntax :CONFigure:AUTO:MODE PROGRAM/0, LOAD/1
Parameter PROGRAM/0, LOAD/1 Auto Load Mode
PROGRAM/0 PROGRAM
LOAD/1 LOAD
Example :CONF:AUTO:MODE 1
Configures Auto Load to
LOAD
Query Syntax :CONFigure:AUTO:MODE?
Return Parameter <NR1> Auto Load Type Status
0 PROGRAM MODE
1 LOAD MODE
Query Example :CONF:AUTO:MODE?
1
Auto load mode is to
LOAD mode.
:CONFigure:SOUND
Channel Specific
Command
Description
Sets the sound of each load module on or off.
Syntax :CONFigure:SOUND OFF/0, ON/1
Parameter OFF/0, ON/1 Sound
OFF/0 Off
ON/1 On
Example :CONF:SOUND ON
Configures the sound for
the specific channel to
On.
Query Syntax :CONFigure:SOUND?
PEL-2000 Programming Manual
38
Return Parameter
<NR1> SOUND Status
0 Off
1 On
Query Example :CONF:SOUND?
0
Sound is off for the
specific channel.
:CONFigure:REMote
All Channels
Description
Turns remote control on or off for all interfaces.
Syntax :CONFigure:REMOTE OFF/0, ON/1
Parameter OFF/0, ON/1 Remote Control
OFF/0 Off
ON/1 On
Example :CONF:REM 1
Turns Remote control
on.
:CONFigure:SAVE
All Channels
Description
This command is for compatibility with other
instruments only and has no action.
Syntax :CONFigure:SAVE
Example :CONF:SAVE
Saves the configuration
data for all channels into
internal memory.
:CONFigure:LOAD
System Command
Description
Configures the load module selector knob as OLD
or Updated.
Syntax :CONFigure:LOAD OLD/0, UPDATED/1