COMMAND DETAILS
103
:SPECification:VOLTage:H/L/C
Channel Specific
Command
Description
Sets the high(H), low(L) and center(C) Go/NoGo
voltage limit specifications. Applicable to CC and
CR mode only.
Syntax :SPECification:VOLTage:H/L/C <NRf+>[V]
Parameter <NRf+>[V] 1 unit = 1 volt Go/NoGo voltage limit
1 1 volt
1V 1 volt
Example :SPEC:VOLT:H 2V
Sets the Go/NoGo high
voltage limit to 2 volts.
Query Syntax :SPECification:VOLTage:H/L/C? <NR2>
Return Parameter <NR2> Go/NoGo voltage
1 unit = 1 volt Returns the limit voltage
Query Example :SPEC:VOLT:H?
2.000
The voltage limit is 2
volts.
:SPECification:CURRent:H/L/C
Channel Specific
Command
Description
Sets the high(H), low(L) and center(C) Go/NoGo
current limit specifications. Applicable to CV
mode only.
Syntax :SPECification:CURRent::H/L/C <NRf+>[A]
Parameter <NRf+>[A] 1 unit = 1 amp Go/NoGo current limit
1 1 amp
1A 1 amp
Example :SPEC:CURR:H 1A
Sets the Go/NoGo high
current limit to 1 amp.
Query Syntax :SPECification:CURRent:H/L/C? <NR2>
PEL-2000 Programming Manual
104
Return Parameter
<NR2> Go/NoGo voltage
1 unit = 1 amp Returns the limit current
Query Example :SPEC:CURR:H?
5.120
The current limit is 5.12
amps.
:SPECification:TEST:
Channel Specific
Command
Description
Turns the Go/NoGo specification (SPEC) limits
on/off.
Syntax :SPECification:TEST OFF/0, ON/1
Parameter OFF/0, ON/1 Go/NoGo limits (SPEC)
OFF/0 OFF
ON/1 ON
Example :SPEC:TEST OFF
Turn Go/NoGo SPEC
off for the specific
channel.
Query Syntax :SPECification:TEST?
Return Parameter <NR1> Go/NoGo SPEC status
0 Off
1 On
Query Example :SPEC:TEST?
1
Go/NoGo SPEC limits is
on.