COMMAND DETAILS
61
Example :CURR:DYNA:LOW:T1 .1S
Sets the T1 time to 100
milliseconds for the
specific channel.
Query Syntax : CURRent:DYNamic:LOW:T1/T2? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] Time
1 Unit=1 second Returns T1/T2 time.
MAX/MIN Returns the allowable
maximum and minimum.
Query Example :CURR:DYN:LOW:T1?
2.5
:CURR:DYN:LOW:T1?
MIN
0.000025
Returns the T1 time of
2.5 seconds.
Returns the minimum T1
time allowable for the
specific channel.
:CURRent:DYNamic:HIGH:L1/L2
Channel Specific
Command
Description
Sets the high range current levels (Level 1 & 2) for
CC dynamic mode.
Syntax :CURRent:DYNamic:HIGH:L1/L2 <NRf+>[A]
Parameter NRf+[A]
L1 10 Sets L1 to 10 Amps. (High
range only)
L2 20 Sets L2 to 20 Amps. (High
Range only)
L1/L2 MIN Sets L1 or L2 to the
minimum level for the
specific channel.
L1/L2 MAX Sets L1 or L2 to the
maximum Level for the
specific channel.
PEL-2000 Programming Manual
62
Example :CURR:DYN:HIGH:L1 10
In high range CC
dynamic mode, Set L1
(level 1) to 10 amps.
Query Syntax :CURRent:DYNamic:HIGH:L1/L2? MIN, MAX
Return Parameter <NR2>[MAX, MIN] Return value
1 unit= 1 amp Returns the current of
Level 1/ 2 (L1/L2).
Query Example :CURR:DYN:HIGH:L2?
MAX
20.4000
Returns the maximum
current allowed for the
channel. (PEL-2020)
:CURRent:DYNamic:HIGH:RISE/FALL
Channel Specific
Command
Description
Sets the high range rising/falling slew rate for CC
dynamic mode for the specific channel.
Syntax :CURRent:DYNamic:HIGH:RISE/FALL <NRf+>[A/uS]
Parameter <NRf+>[A/uS] Slew Rate
RISE/FALL 0.8A/uS Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL 1 Sets the rising/falling slew
rate to 1A/uS
RISE/FALL MIN Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX Sets to the fastest
rising/falling slew rate.
Example :CURR:DYNA:HIGH:RISE
1.1
Sets the rising slew rate
to 1.1A/uS.
Query Syntax : CURRent:DYNamic:HIGH:RISE/FALL? [MIN, MAX]
Return Parameter <NR2>[MAX|MIN] Slew rate
1 Unit = 1 amp/uS Returns the slew rate.
MAX/MIN Returns the allowable
maximum and minimum.