COMMAND DETAILS
39
Example :CONF:LOAD UPDATED
Sets the load module
selector knob as
Updated.
Parameter OLD/0, UPDATED/1 Configuration type
OLD/0 Old
UPDATED/1 Updated
Example :CONF:LOAD OLD
Configuration type set as
OLD.
Query Syntax :CONFigure:LOAD?
Return Parameter <NR1> Configuration type
0 Old
1 Updated
Query Example :CONF:LOAD?
0
Sets the load module
selector configuration
type as OLD.
:CONFigure:PROTection:CURRent:STATe
Channel Specific
Command
Description
Sets the current protection for the specific channel
on or off. The current protection can also be
cleared.
Syntax :CONFigure:PROTection:CURRent:STATe OFF/0,
ON/1, CLEAR/2
Parameter OFF/0, ON/1, CLEAR/2 Current Protection
CLEAR/2 Cleared
OFF/0 Off
ON/1 On
Example
:CONF:PROT:CURR:STAT 1
Turns on current protection.
Query Syntax : CONFigure:PROTection:CURRent:STATe?
Return Parameter <NR1> Current Protection
0 Off
PEL-2000 Programming Manual
40
1 On
2 Clear
Query Example
:CONF:PROT:CURR:STAT?
1
Current protection is
turned on.
:CONFigure:PROTection:CURRent:LEVel
Channel Specific
Command
Description
Sets the current protection level for the
current/specific channel. The level can be set to
any applicable level or to the channel
maximum/minimum.
Syntax :CONFigure:PROTection:CURRent:LEVel <NRf>[A],
MIN, MAX
Parameter <NRf> Current Protection Level
.3 300mA
0.3A 300mA
300MA 300mA
MIN Sets to the minimum level
MAX Sets the current limit to
the maximum level
Example
:CONF:PROT:CURR:LEV
MAX
Sets the current limit to
20.40A (PEL2020)
Query Syntax : CONFigure:PROTection:CURRent:LEVel?
Return Parameter <NRf> 1 unit = 1 amp Current protection level
1 1 amp.
Query Example
:CONF:PROT:CURR:LEV?
0.30
Current protection level
is at 300mA.