COMMAND DETAILS
41
:CONFigure:PROTection:VOLTage:STATe
Channel Specific
Command
Description
Sets the voltage protection for the current/specific
channel on or off. The voltage protection can also
be cleared.
Syntax :CONFigure:PROTection:VOLTage:STATe OFF/0,
ON/1, CLEAR/2
Parameter OFF/0, ON/1, CLEAR/2 Voltage Protection
CLEAR/2 Clear
OFF/0 Off
ON/1 On
Example
:CONF:PROT:VOLT:STAT 1
Turns on voltage
protection.
Query Syntax : CONFigure:PROTection:VOLTage:STATe?
Return Parameter <NR1> Voltage Protection state
0 Off
1 On
2 Clear
Query Example
:CONF:PROT:VOLT:STAT?
0
Voltage protection is
currently off.
:CONFigure:PROTection:VOLTage:LEVel
Channel Specific
Command
Description
Sets the voltage protection level for the
current/specific channel. The level can be set to
any applicable level or to the channel
maximum/minimum.
Syntax :CONFigure:PROTection:VOLTage:LEVel <NRf>[V],
MIN, MAX
Parameter <NRf> Voltage Protection Level
30 30 volts
PEL-2000 Programming Manual
42
30V 30 volts
MIN Sets to the minimum level
MAX Sets the voltage limit to
the maximum level
Example
:CONF:PROT:VOLT:LEV MAX
Sets the voltage limit to
81.6V (PEL2020)
Query Syntax : CONFigure:PROTection:VOLTage:LEVel?
Return Parameter <NRf> 1 unit = 1 volt Voltage protection level
1.00 1.00 volts.
Query Example
:CONF:PROT:VOLT:LEV?
81.6000
Voltage protection level
is at 81.6V.
:CONFigure:PROTection:POWer:STATe
Channel Specific
Command
Description
Sets the power protection for the current/specific
channel on or off. The power protection can also be
cleared.
Syntax :CONFigure:PROTection:POWer:STATe OFF/0, ON/1,
CLEAR/2
Parameter OFF/0, ON/1, CLEAR/2 Power Protection
CLEAR/2 Cleared
OFF/0 Off
ON/1 On
Example
:CONF:PROT:POW:STAT 1
Turns on power protection.
Query Syntax : CONFigure:PROTection:POWer:STATe?
Return Parameter <NR1> Power Protection
0 Off
1 On
2 Clear
Query Example
:CONF:PROT:POW:STAT?
1
Power protection is
currently on.