COMMAND DETAILS
57
:CURRent:DYNamic:RISE/FALL
Channel Specific
Command
Description
Sets the rising/falling slew rate for CC dynamic
mode for the specific channel and range.
Syntax :CURRent:DYNamic:RISE/FALL <NRf+>[A/uS]
Parameter <NRf+>[A/uS] Slew Rate
RISE/FALL 0.8A/uS Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL 1 Sets the rising/falling slew
rate to 1A/uS
RISE/FALL MIN Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX Sets to the fastest
rising/falling slew rate.
Example :CURR:DYNA:RISE 1.1
Sets the rising slew rate
to 1.1A/uS.
Query Syntax : CURRent:DYNamic:RISE/FALL? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] Slew rate
1 Unit=1 amp/uS Returns the slew rate.
MAX/MIN Returns the allowable
maximum or minimum.
Query Example :CURR:DYN:FALL? MIN
0.0003
Shows the minimum
allowable value for the
falling slew rate as 0.0003
A/uS for the specific
channel and range.
:CURRent:DYNamic:T1/T2
Channel Specific
Command
Description
Sets the timers T1 or T2 for CC dynamic mode for
the specific channel and range.
Syntax :CURRent:DYNamic:T1/T2 <NRf+>[S|ms]
PEL-2000 Programming Manual
58
Parameter
<NRf+>[S] Time
T1/T2 0.1S Sets the T1/T2 time to 0.1
seconds.
T1/T2 1 Sets T1/T2 to 1 second.
T1/T2 MIN Sets the T1/T2 to the
minimum value.
T1/T2 MAX Sets the T1/T2 time to the
maximum time
Example :CURR:DYNA:T1 .1S
Sets the T1 time to 100
milliseconds for the
specific channel.
Query Syntax : CURRent:DYNamic:T1/T2? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] Time
1 Unit=1 second Returns T1/T2 time.
MAX/MIN Returns the allowable
maximum and minimum.
Query Example :CURR:DYN:LOW:T1?
2.5
:CURR:DYN:LOW:T1?
MIN
0.000025
Returns the T1 time of
2.5 seconds.
Returns the minimum T1
time allowable for the
specific channel and
range.
:CURRent:DYNamic:LOW:L1/L2
Channel Specific
Command
Description
Sets the low range current levels (Level 1 & 2) for
CC dynamic mode.
Syntax :CURRent:DYNamic:LOW:L1/L2 <NRf+>[A]
Parameter NRf+[A] Current
L1 1 Sets L1 to 1 Amp. (low
range only)