COMMAND DETAILS
91
:RESistance:STATic:HIGH:AVALue/BVALue
Channel Specific
Command
Description
Sets the high range A/B Value for constant
resistance static mode.
Syntax :RESistance:STATic:HIGH:AVALue/BVALue
<NRf+>[OHM]
Parameter NRf+[OHM]
AVALue 10 Sets A Value to 10 ohms.
(high range only)
BVALue 20OHM Sets B Value to 20 ohms.
(high Range only)
AVALue MIN Sets A Value to the
minimum level for the
specific channel.
AVALue MAX Sets A Value to the
maximum level for the
specific channel.
Example :RES:STAT:HIGH:BVAL 10
Sets high range CR static
mode B Value to 10
ohms.
Query Syntax :RESistance:STATic:HIGH:AVALue/BVALue?
Return Parameter <NR2>[MAX, MIN] Resistance
1 unit= 1 ohm Returns the resistance of
the A or B Value.
Query Example :RES:STAT:HIGH:BVAL?
MAX
15000.0
Returns the maximum
resistance allowed for
the channel for B Value.
(PEL-2020)
:RESistance:STATic:HIGH:RISE/FALL
Channel Specific
Command
Description
Sets the high range rising/falling slew rate.
Syntax :RESistance:STATic:HIGH:RISE/FALL <NRf+>[A/uS]
PEL-2000 Programming Manual
92
Parameter
<NRf+>[A/uS] Slew rate
RISE/FALL 0.8A/uS Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL 0..5 Sets the rising/falling slew
rate to 0.5A/uS
RISE/FALL MIN Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX Sets to the fastest
rising/falling slew rate.
Example :RES:STAT:HIGH:RISE 1.1
Sets the rising slew rate
to 1.1A/uS.
Query Syntax :RESistance:STATic:HIGH:RISE/FALL? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] Slew rate
1 Unit=1 amp/uS Returns the slew rate.
MAX/MIN Returns the allowable
maximum and minimum.
Query Example :RES:STAT:HIGH:RISE?
MIN
0.8000
For high range CR mode,
the minimum value for
the rising slew rate is 0.8
A/uS for the specific
channel.
:RESistance:DYNamic:LOW:L1/L2
Channel Specific
Command
Description
Sets the low range resitance levels (Level 1 & 2) for
CR dynamic mode.
Syntax :RESistance:DYNamic:LOW:L1/L2 <NRf+>[OHM]
Parameter NRf+[OHM] Ohms
L1 10 Sets L1 to 10 ohms. (low
range only)
L2 20OHM Sets L2 to 20 ohms. (low
range only)