EasyManua.ls Logo

GW Instek PEL-2002 - RUN Subsystem; Resistance:dynamic:high:t1;T2; Run

GW Instek PEL-2002
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND DETAILS
97
:RESistance:DYNamic:HIGH:T1/T2
Channel Specific
Command
Description
Sets the timers T1 and T2 for high range dynamic
CR mode.
Syntax :RESistance:DYNamic:HIGH:T1/T2 <NRf+>[S|ms]
Parameter <NRf+>[S] Timer T1/T2
T1/T2 0.1S Sets the T1/T2 time to 0.1
seconds.
T1/T2 1 Sets T1/T2 to 1 second.
T1/T2 MIN Sets the T1/T2 to the
minimum value.
T1/T2 MAX Sets the T1/T2 time to the
maximum time
Example :RES:DYNA:HIGH:T1 10S
Sets the high range T1
time to 10 seconds for
the specific channel.
Query Syntax :RESistance:DYNamic:HIGH:T1/T2? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] T1/T2 time.
1 Unit=1 second Returns T1/T2 time.
MAX/MIN Returns the allowable
maximum and minimum.
Query Example :RES:DYN:HIGH:T1?
2.5
:RES:DYN:LOW:T1? MIN
0.000025
Returns the T1 time of
2.5 seconds.
Returns the minimum T1
time allowable for the
specific channel.
PEL-2000 Programming Manual
98
RUN Subsystem
:RUN ............................................................................ 98
:RUN
All Channel
Command
Description
Turns on all the electronic loads.
Syntax :RUN
Example :RUN
Turns on all electronic