COMMAND DETAILS
87
Resistance Subsystem
:RESistance:L1/L2 ....................................................... 87
:RESistance:RISE/FALL ............................................... 88
:RESistance:STATic:LOW:AVALue/BVALue ............... 89
:RESistance:STATic:LOW:RISE/FALL ......................... 90
:RESistance:STATic:HIGH:AVALue/BVALue ............. 91
:RESistance:STATic:HIGH:RISE/FALL ....................... 91
:RESistance:DYNamic:LOW:L1/L2 ............................ 92
:RESistance:DYNamic:LOW:RISE/FALL .................... 93
:RESistance:DYNamic:LOW:T1/T2 ............................ 94
:RESistance:DYNamic:HIGH:L1/L2 ........................... 95
:RESistance:DYNamic:HIGH:RISE/FALL................... 96
:RESistance:DYNamic:HIGH:T1/T2 .......................... 97
:RESistance:L1/L2
Channel Specific
Command
Description
Sets A/B Value for constant resistance mode,
where L1 is A Value and L2 is B Value. This
command only applies to current mode (static).
Syntax :RESistance:L1/L2 <NRf+>[OHM]
Parameter NRf+[OHM] Resistance
L1 10 Sets A Value to 10 ohms
L2 20 Sets B Value* to 20 ohms
*Single Channel
L1 MIN Sets A Value to the
minimum level for the
specific channel.
L1 MAX Sets A Value to the
maximum level for the
specific channel.
PEL-2000 Programming Manual
88
Example :RES:L1 10
Sets CR static mode A
Value to 10 ohms,
depending on the
specific range
Query Syntax :RESistance:L1/L2?
Return Parameter <NR2>[MAX, MIN] Resistance
1 unit = 1 ohm/1 k ohm Returns the resistance of
the A or B Value.
Query Example :RES:L1? MAX
300
Returns the maximum
resistance allowed for
the channel. (PEL-2020)
:RESistance:RISE/FALL
Channel Specific
Command
Description
Sets the rising/falling slew rate for CR mode. The
command applies to the current mode
(static/dynamic) and the current range
(High/Low)
Syntax :RESistance::RISE/FALL <NRf+>[A/uS]
Parameter <NRf+>[A/uS] Slew rate
RISE/FALL 0.8A/uS Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL .8 Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL MIN Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX Sets to the fastest
rising/falling slew rate.
Example :RES:RISE 0.1
Sets the rising slew rate
to 0.1A/uS.
Query Syntax :RESistance:RISE/FALL? [MIN|MAX]