EasyManua.ls Logo

GW Instek PEL-2002 - Resistance:dynamic:low:rise;Fall; Resistance:dynamic:low:t1;T2

GW Instek PEL-2002
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND DETAILS
93
L1/L2 MIN Sets L1 or L2 to the
minimum level for the
specific channel.
L1/L2 MAX Sets L1 or L2 to the
maximum Level for the
specific channel.
Example :RES:DYN:LOW:L1 10
In low range CR
dynamic mode, Set L1
(level 1) to 10 ohms.
Query Syntax :RESistance:DYNamic:LOW:L1/L2? MIN, MAX
Return Parameter <NR2>[MAX, MIN] Resistance
1 unit= 1 ohm Returns the resistance of
L1/L2.
Query Example :RES:DYN:LOW:L2? MAX
300
Returns the maximum
resistance allowed for
the channel. (PEL-2020)
:RESistance:DYNamic:LOW:RISE/FALL
Channel Specific
Command
Description
Sets the low range rising/falling slew rate for CR
dynamic mode for the specific channel.
Syntax :RESistance:DYNamic:LOW:RISE/FALL <NRf+>[A/uS]
Parameter <NRf+>[A/uS] Slew Rate
RISE/FALL 0.8A/uS Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL .1 Sets the rising/falling slew
rate to 0.1A/uS
RISE/FALL MIN Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX Sets to the fastest
rising/falling slew rate.
Example :RES:DYNA:LOW:RISE 0.1
Sets the rising slew rate
to ~ 0.1A/uS.
PEL-2000 Programming Manual
94
Query Syntax :RESistance:DYNamic:LOW:RISE/FALL? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] Slew rate
1 Unit=1 amp/uS Returns the slew rate.
MAX/MIN Returns the allowable
maximum and minimum.
Query Example :RES:DYN:LOW:FALL?
MIN
0.8000
For low range dynamic
CR mode, the minimum
value for the falling slew
rate is 0.8 A/uS for the
specific channel.
:RESistance:DYNamic:LOW:T1/T2
Channel Specific
Command
Description
Sets the timers T1 or T2 for CR dynamic mode for
the specific channel in low range.
Syntax :RESistance:DYNamic:LOW:T1/T2 <NRf+>[S|ms]
Parameter <NRf+>[S] Time T1/T2
T1/T2 0.1S Sets the T1/T2 time to 0.1
seconds.
T1/T2 1 Sets T1/T2 to 1 second.
T1/T2 MIN Sets the T1/T2 to the
minimum value.
T1/T2 MAX Sets the T1/T2 time to the
maximum time
Example :RES:DYNA:LOW:T1 10S
Sets the T1 time to 10
seconds for the specific
channel.
Query Syntax :RESistance:DYNamic:T1/T2? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] Time T1/T2
1 Unit=1 second Returns T1/T2 time.
MAX/MIN Returns the allowable
maximum and minimum.

Table of Contents

Other manuals for GW Instek PEL-2002

Related product manuals