COMMAND DETAILS
85
Example :PROG:ONT 10S
Set the on-time for the
current program number
to 10 seconds.
Query Syntax :PROGram:ONTime? <NR2>
Return Parameter
<NR2>
Program On Time
0.1-60 0.1~60 seconds
Query Example :PROG:ONT?
10
Returns the on-time for
the current program
number in seconds.
:PROGram:OFFTime
Program Number
Specific
Description
Sets the off-time for the program number. 0.1~60
seconds.
Syntax :PROGram:OFFTime <NRf>[S]
Parameter
<NRf>[S]
Program Off Time
0.1-60 0.1~60 seconds
0.1-60s 0.1~60 seconds
Example :PROG:OFFT 10S
Set the off-time for the
current program number
to 10 seconds.
Query Syntax :PROGram:OFFTime? <NR2>
Return Parameter
<NR2> I unit = 1 second Program Off Time
0.1-60 0.1~60 seconds
Query Example :PROG:OFFT?
10
Returns the off-time for
the current program
number in seconds.
PEL-2000 Programming Manual
86
:PROGram:RUN
All Channel
Command
Description
Runs the current program number when set to on,
and when set to off will allow all the
program/sequence data to be programmed.
Syntax :PROGram:RUN [OFF/0, ON/1]
Parameter
OFF/0, ON/1 Run Program
OFF/0 OFF
ON/1 ON
Example :PROG:RUN 1
Runs the program.
:PROGram:SAVE
All Channel
Command
Description
Saves the current program to memory.
Syntax :PROGram:SAVE
Example :PROG:SAVE
Saves the program to
memory.