COMMAND DETAILS
65
:FETCh:POWer?
Channel Specific
Status Command
Description
This query returns the real-time power of the load
module input for the specific channel.
Syntax :FETCh:CURRent? <NR2>
Parameter <NR2> 1 unit= 1 amp
1 1 amp
Query Example :FETC:POW?
1.2
The specific channel is at
1.2 watts.
:FETCh:STATus?
Status Command
Description
This query returns the status of the load module.
The returned value is the bit weight of the Channel
Status Register. See page 131.
Syntax :FETCh:STATus? <NR1>
Parameter <NR1> Condition <NR1> Condition
1 OC 16 OT
2 OV 32 G/N
4 OP 64 UVP
8 RV 16-128 Not Used
Query Example :FETC:STAT?
2
Over voltage (OV)
protection has been
triggered for the specific
channel.
:FETCh:ALLVoltage?
All Channel Status
Command
Description
This query returns the voltage values of all the
load modules/channels in order from 1-8(PEL-
2004)/1-4(PEL2002).
Syntax :FETCh:ALLVoltage?
PEL-2000 Programming Manual
66
Parameter
<aard>
CH1,CH2,CH3,CH4,CH5,
CH6,CH7,CH8
Returns all the voltage
values from all the
channels, 1-8(PEL-
2004)/1-4(PEL-2002).
Query Example :FETC:ALLV?
2.5000, 3.0000, 0.0000,
0.0000, 0.0000, 0.0000,
5.500, 0.0000
Channel 1 and 2 have
voltages of 2.5 and 3
volts respectively.
Channels 3-6 and 8 have
no voltage and channel
7 is 5.5 volts
:FETCh:ALLCurrent?
All Channel Status
Command
Description
This query returns the current values of all the
load modules/channels in order from 1-8(PEL-
2004)/1-4(PEL2002).
Syntax :FETCh:ALLCurrent? <aard>
Parameter
<aard>
CH1,CH2,CH3,CH4,CH5,
CH6,CH7,CH8
Returns all the current
values from all the
channels, 1-8(PEL-
2004)/1-4(PEL-2002).
Query Example :FETC:ALLC?
0.0000, 0.0000, 0.0000,
0.0000, 0.0000, 0.0000,
1.2000, 3.5600
Channels 1 to 6 have no
current. Channels 7 & 8
have 1.2 and 3.56 amps,
respectively.
:FETCh:ALLPower?
All Channel Status
Command
Description
This query returns the power values of all the load
modules/channels in order from 1-8(PEL-2004)/1-
4(PEL2002).
Syntax :FETCh:ALLPower? <aard>