COMMAND DETAILS
43
:CONFigure:PROTection:POWer:LEVel
Channel Specific
Command
Description
Sets the power protection level for the
current/specific channel. The level can be set to
any applicable level or to the channel
maximum/minimum.
Syntax :CONFigure:PROTection:POWer:LEVel <NRf>[W],
MIN, MAX
Parameter <NRf> Power Protection Level
200 200Watts
200W 200Watts
MIN Sets to the minimum level
MAX Sets the power limit to the
maximum level
Example
:CONF:PROT:POW:LEV MAX
Sets the power limit to
102W (PEL2020)
Query Syntax : CONFigure:PROTection:POWEr:LEVel?
Return Parameter <NRf> Power protection level
1 unit = 1 watt Returns the power
protection level in Watts.
Query Example
:CONF:PROT:POW:LEV?
75
Power protection level is
at 75 watts.
:CONFigure:PROTection:UVP:CLEar
All Channel
Command
Description
Clears the under voltage power protection status.
Syntax :CONFigure:PROTection:UVP:CLEar
Example :CONF:PROT:UVP:CLE
Clears the under voltage
protection.
PEL-2000 Programming Manual
44
:CONFigure:PROTection:UVP:LEVel
Channel Specific
Command
Description
Sets the under voltage protection level for the
current/specific channel. The level can be set to
any applicable level or to the channel
maximum/minimum.
Syntax :CONFigure:PROTection:UVP:LEVel <NRf>[W], MIN,
MAX
Parameter <NRf> Under Voltage Protection
Level
20 20 Volts
20V 20 Volts
MIN Sets to the minimum level
(OFF)
MAX Sets the voltage limit to
the maximum level
Example
:CONF:PROT:UVP:LEV MIN
Sets the UVP limit to OFF
Query Syntax : CONFigure:PROTection:UVP:LEVel?
Return Parameter <NRf> Power protection level
1 unit = 1 volt Returns the UVP level as
volts.
Query Example
:CONF:PROT:UVP:LEV?
75
UVP level is at 75 volts.