COMMAND DETAILS
63
Query Example :CURR:DYN:HIGH:FALL?
MAX
0.8
For high range dynamic
CC mode, the maximum
value for the falling slew
rate is 0.8 A/uS for the
specific channel.
:CURRent:DYNamic:HIGH:T1/T2
Channel Specific
Command
Description
Sets the timers T1 or T2 for CC dynamic mode for
the specific channel in high range.
Syntax :CURRent:DYNamic:HIGH:T1/T2 <NRf+>[S|ms]
Parameter <NRf+>[S] Time
T1/T2 0.1S Sets the T1/T2 time to 0.1
seconds.
T1/T2 1 Sets T1/T2 to 1 second.
T1/T2 MIN Sets the T1/T2 to the
minimum value.
T1/T2 MAX Sets the T1/T2 time to the
maximum time
Example :CURR:DYNA:HIGH:T1
10S
Sets the high range T1
time to 10 seconds for
the specific channel.
Query Syntax : CURRent:DYNamic:HIGH:T1/T2? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] Time
1 Unit=1 second Returns T1/T2 time.
MAX/MIN Returns the allowable
maximum and minimum.
Query Example :CURR:DYN:HIGH:T1?
2.5
:CURR:DYN:HIGH:T1?
MIN
0.000025
Returns the T1 time of
2.5 seconds.
Returns the minimum T1
time allowable for the
specific channel.
PEL-2000 Programming Manual
64
FETCH Subsystem
:FETCh:VOLTage? ........................................................ 64
:FETCh:CURRent? ........................................................ 64
:FETCh:POWer? ........................................................... 65
:FETCh:STATus? .......................................................... 65
:FETCh:ALLVoltage? .................................................... 65
:FETCh:ALLCurrent? .................................................... 66
:FETCh:ALLPower?....................................................... 66
:FETCh:VOLTage?
Channel Specific
Status Command
Description
This query returns the real-time voltage of the load
module input for the specific channel.
Syntax :FETCh:VOLTage? <NR2>
Parameter <NR2> 1 unit = 1 volt Voltage
8 8 volts
Query Example :FETC:VOLT?
11.2
The specific channel has
a voltage of 11.2 volts at
the input.
:FETCh:CURRent?
Channel Specific
Status Command
Description
This query returns the real-time current of the load
module input for the specific channel.
Syntax :FETCh:CURRent? <NR2>
Parameter <NR2> 1 unit= 1 amp
1 1 amp
Query Example :FETC:CURR?
1.2
The specific channel has
a current of 1.2 amps at
the load module input.