COMMAND DETAILS
73
:MEASure:POWer?
Channel Specific
Command
Description
This query returns the measured power of the
specific channel.
Query Syntax :MEASure:POWer? <NR2>
Return Parameter <NR2> 1 unit = 1 watt Power at the load input
1.0000 1.0000 watts
Query Example :MEAS:POW?
1.5
1.5 watts is measured at
the specific channel load
input.
:MEASure:INPut
Channel Specific
Command
Description
This command is for compatibility with other
instruments only and has no action.
Syntax :MEASure:INPut [LOAD/0, UUT/1]
Parameter LOAD/0, UUT/1 Voltage Sense
LOAD/0 Disabled
UUT/1 Enabled
Example :MEAS:INP 0
Disable voltage sense.
Query Syntax :MEASure:INPut? <NR1>
Return Parameter <NR1> Voltage Sense
0 Disabled
1 Enabled
Query Example :MEAS:INP?
1
Returns the voltage
input status. Voltage
sense is enabled.
PEL-2000 Programming Manual
74
:MEASure:SCAN
Channel Specific
Command
Description
This command allows the mainframe to scan all
the load modules.
Syntax :MEASure:SCAN [OFF/0, ON/1]
Parameter OFF/0, ON/1 Scan
OFF/0 Disabled
ON/1 Enabled
Example :MEAS:SCAN 0
Disable scanning.
Query Syntax :MEASure:SCAN? <NR1>
Return Parameter <NR1> Scan
0 Disabled
1 Enabled
Query Example :MEAS:SCAN?
1
Returns the scanning
status. Here scanning is
enabled.
:MEASure:ALLVoltage?
All Channel
Command
Description
This query measures the voltage values of all the
load modules/channels in order from 1-8 (PEL-
2004)/1-4(PEL2002).
Query Syntax :MEASure:ALLVoltage? <aard>
Query Parameter
<aard> 1 unit = 1 volt
CH1,CH2,CH3,CH4,CH5,
CH6,CH7,CH8
Returns all the voltage
values from all the
channels, 1-8(PEL-
2004)/1-4(PEL-2002).