COMMAND DETAILS
129
Return Parameter <NR1> Point
1~max channels Returns the channel
duration time that the
specific channel is using.
Query Example :SEQ:CHAN:TIME?
2
The specific channel is
using the channel
duration setting of
channel 2.
:SEQuence:RUN
Channel Specific
Command
Description
Turns the sequence On/Off.
Syntax :SEQuence:RUN OFF/0, ON/1
Parameter OFF/0, ON/1 Sequence
OFF/0 Off
ON/1 On
Example :SEQ:RUN ON
Run the sequence.
:SEQuence:SAVE
Channel Specific
Command
Description
Saves the sequence for the specific channel.
Syntax :SEQuence:SAVE
Example :SEQ:SAVE
Saves the sequence.
PEL-2000 Programming Manual
130
Command Error Codes
Description
The PEL has a number of specific error codes. Use
the SYSTem:ERRor command to recall the error
codes.
-102 Syntax error.
An unrecognized command or data
type was encountered.
-109 Missing parameter
The command header requires
more parameters than was received.
-122 Data out of range
The data is outside the allowed
range.
-128 Numeric data not allowed
The command does not accept
numerical data/parameters
-200 Execution error
Generic execution error.
-144 Character Data too long
The character data contains more
than twelve characters
-151 Invalid String
The string data received is invalid
-148 Character data not allowed
The command does not accept
character data
-138 Suffix not allowed
A command does not accept
suffixes/the suffix type.