COMMAND DETAILS
127
Return Parameter <NR1> On End Load
0 Off
1 On
Query Example :SEQ:END:LOAD?
1
On End Load is On.
:SEQuence:VOLTage:RANGe
Channel Specific
Command
Description
Sets the sequence CC voltage range.
Syntax :SEQuence:VOLTage:RANGe <NRf>[V], L, H
Parameter <NRf>[ V] , L, H Range
16 Low range*
80V High range*
L Low range
H High range
*Load module dependent, PEL-2020 shown.
Example :SEQ:VOLT:RANG L
Sets the range to Low for
the channel.
Query Syntax :SEQuence:VOLTage:RANGe?
Return Parameter <NR2> Range
16 Low PEL-2020,2030,2040
125 Low PEL-2041
80 High PEL-2020,2030,2040
500 High PEL-2041
Query Example :SEQ:VOLT:RANG?
500
Returns the voltage
range. In this case high
for the PEL-2041.
PEL-2000 Programming Manual
128
:SEQuence:LOOP:STARt
Channel Specific
Command
Description
Determines from which point to start repeating
(looping) the sequence from when using the
SEQuence:REPeat command.
Syntax :SEQuence:LOOP:STARt <NR1>
Parameter <NR1> Start loop from
1~last point 1st~ last point.
Example :SEQ:LOOP:STAR 2
Loop from point 2.
Query Syntax :SEQuence:LOOP:STARt?
Return Parameter <NR1> Point
1~last point Returns the point that the
loop will start from.
Query Example :SEQ:LOOP:STAR?
2
The loop starts at point
2.
:SEQuence:CHANnel:TIME
Channel Specific
Command
Description
Sets which channel duration time the specific
channel will use.
Syntax :SEQuence:CHANnel:TIME <NR1>
Parameter <NR1>1~max channels Duration Time Settings
1 Use channel 1
2 etc Use channel 2 etc
Example :SEQ:CHAN:TIME 3
Set the specific channel
to use the channel
duration time of channel
3.
Query Syntax :SEQuence:CHANnel:TIME?