COMMAND DETAILS
51
L1 MIN Sets A Value to the
minimum level for the
specific channel.
L1 MAX Sets A Value to the
maximum Level for the
specific channel.
Example :CURR:STAT:L1 1
Sets A Value to 1 amp
for the current range
Query Syntax :CURRent:STATic:L1?/L2? [MAX, MIN]
Return Parameter <NR2>[MAX, MIN] Current
1 unit = 1 amp Returns the current of the
A Value (L1) or B Value
(L2).
Query Example :CURR:STAT:L2? MAX
10.2
Returns the maximum
current allowed for the
channel. (PEL-2020)
Query Example :CURR:STAT:L2? 2
Returns the current
setting (2 A) for B Value.
:CURRent:STATic:RISE/FALL
Channel Specific
Command
Description
Sets the slew rate for constant current static mode.
The command is range dependant. If the current
range is Low, then the command will only apply to
the low range settings.
Syntax :CURRent:STATic:RISE/FALL <NRf+>[A/uS]
Parameter <NRf+>[A/uS] Slew Rate
RISE/FALL 0.078A/uS Sets the rising/falling slew
rate to 0.078A/uS
RISE/FALL 1 Sets the rising/falling slew
rate to 1A/uS
PEL-2000 Programming Manual
52
RISE/FALL MIN Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX Sets to the fastest
rising/falling slew rate.
Example
:CURR:STAT:RISE .01
Sets the rising slew rate
to 0.01A/uS.
Query Syntax : CURRent:STATic:RISE/FALL? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] Slew rate
1 Unit=1 amp/uS Returns the slew rate.
MAX/MIN Returns the allowable
maximum and minimum.
Query Example :CURR:STAT:RISE? MIN
0.078
the Minimum value for
the rising slew rate is
0.078 A/uS for the
specific channel.
Query Example :CURR:STAT:RISE?
0.16800
The rising slew rate is
0.168 A/uS for the
specific channel.
:CURRent:STATic:LOW:AVALue/BVALue
Channel Specific
Command
Description
Sets the low range A/B Value for constant current
static mode.
Syntax :CURRent:STATic:LOW:AVALue/BVALue <NRf+>[A]
Parameter NRf+[A]
AVALue 1 Sets A Value to 1 Amp.
(Low range only)
BVALue 2 Sets B Value to 2 Amps.
(Low Range only)
AVALue 1A Sets A Value to 1 Amp.
(Low range only)