COMMAND DETAILS
89
Return Parameter <NR2>[MAX|MIN] Slew rate
1 Unit =1 amp/uS Returns the slew rate.
MAX/MIN Returns the allowable
maximum and minimum.
Query Example :RES:RISE? MAX
0.8000
Returns the maximum
value for the rising slew
rate (0.8 A/uS).
:RESistance:STATic:LOW:AVALue/BVALue
Channel Specific
Command
Description
Sets the low range A/B Value for constant
resistance static mode.
Syntax :RESistance:STATic:LOW:AVALue/BVALue
<NRf+>[OHM]
Parameter NRf+[OHM] Resistance
AVALue 10 Sets A Value to 10 ohms.
(Low range only)
BVALue 20 Sets B Value to 20 ohms.
(Low Range only)
AVALue MIN Sets A Value to the
minimum level for the
specific channel.
AVALue MAX Sets A Value to the
maximum level for the
specific channel.
Example :RES:STAT:LOW:BVAL 10
Sets low range CR static
mode B Value to 10
ohms.
Query Syntax :RESistance:STATic:LOW:AVALue/BVALue?
Return Parameter <NR2>[MAX, MIN] Resistance
1 unit = 1 ohm Returns the resistance of
the A or B Value.
PEL-2000 Programming Manual
90
Query Example :RES:STAT:LOW:AVAL?
MAX
300
Returns the maximum
resistance allowed for
the channel. (PEL-2020)
:RESistance:STATic:LOW:RISE/FALL
Channel Specific
Command
Description
Sets the low range rising/falling slew rate.
Syntax :RESistance:STATic:LOW:RISE/FALL <NRf+>[A/uS]
Parameter <NRf+>[A/uS] Slew rate
RISE/FALL 0.8A/uS Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL .8 Sets the rising/falling slew
rate to 0.8 A/uS
RISE/FALL MIN Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX Sets to the fastest
rising/falling slew rate.
Example :RES:STAT:LOW:RISE 0.1
Sets the rising slew rate
to 0.1A/uS.
Query Syntax :RESistance:STATic:LOW:RISE/FALL? [MIN, MAX]
Return Parameter <NR2>[MAX, MIN] Slew rate
1 Unit =1 amp/uS Returns the slew rate.
MAX, MIN Returns the allowable
maximum and minimum.
Query Example :RES:STAT:LOW:RISE?
MAX
0.8000
For low range CR mode,
the maximum value for
the rising slew rate is 0.8
A/uS for the specific
channel.